or call +91 91766 77448

Altitude Test Chamber

For Combined Altitude with Temperature & Humidity Simulations

Altitude Test Chamber

KLIMA AT range of Altitude Test Chambers seamlessly combine temperature and altitude settings, with the added option of humidity control, allowing for simultaneous environmental testing. Users can easily adjust the chamber temperature while subjecting the product under test to varying altitude levels, reaching up to 100mBar. This innovative blend of features enables precise simulation of real-life conditions, providing invaluable insights into how products perform at different altitude levels.

  • Temperature Range: -70ËšC to +180ËšC (Cascade) & -40ËšC to +180ËšC (Single)
  • Altitude Range: AP to 100mBar
  • Humidity Range: 10% RH to 98% RH
  • Rapid Decompression in controlled environment
  • Durable welded pressure chamber for structural strength
  • Slide-Out, Adjustable Product Shelf
  • Lockable, One-Handed Door Latch
  • Durable Silicone Door Gasket
  • Two 4″ (100mm) Circular Port Plugs
  • TCP/IP and Serial Communication
  • Frost-Free Operation
  • Non-Settling Rigid Stone Wool Insulation
  • Heavy-Duty Swivel Casters with Leveling Pads
  • Energy Saving Refrigerants R404A and R23
  • Fog-Free Tempered Heated Viewing Window with Interior Light
  • Continuous TIG Welding 304 Stainless Steel Interior Liner
  • Easy Maintenance with Three-Sided Access to Components

Simulate the Summit: Altitude Testing Made Easy.

KLIMA AT is customizable to conduct testing as per your choice. Below are some of the frequently used test standards for Altitude Testing

IEC 60068 - 2 - 30, 2005, Environmental testing Part 2-30:, Tests: Damp heat, cyclic
IEC 60068 - 2 - 78, 2012, Environmental testing Part 2-78:, Tests: Damp heat, steady state
IEC 60068 - 2 - 1, 2007, Environmental testing Part 2-1:, Tests: Cold
IEC 60068 - 2 - 2, 2007, Environmental testing Part 2-2:, Tests: Dry heat
IEC 60068 - 2 - 14, 2009, Environmental testing Part 2-14: Tests: Change of temperature
IEC 60068-2-131
ASTM D6693
MIL-STD-810H

Technical Specifications

 Specification/ Model
KLIMA-216/-60C/5KAT
TEST SPACE DIMENSIONS
Test space volume Ltr. 216
Width mm 600
Depth mm 600
Height mm 600
PERFORMANCE DATA FOR TEMPERATURE 
Controllable AP mbar 100
Maximum temperature °C 100
Minimum temperature discontinuous °C -60
Rate of temperature change, cooling  °C/min 5
Rate of temperature change, heating °C/min 5
Control Accuracy, in time °C ±1
Heat compensation, max. at +20 °C W 850
PERFORMANCE DATA FOR VACCUM
Pumping Range mbar AP to 100
Pumping time to 10mbar min <30
pressure stability mbar ±3
PERFORMANCE DATA FOR HUMIDITY
Maximum temperature °C 85
Minimum temperature  °C 10
Humidity range  %RH 10…98
Humidity deviation, in time %RH ±1…±3
Temperature deviation, in time %RH ±1
Test Standard Compliance IEC-60068, IEC-60749, IEC-61747, MIL-STD, ISO16750, ASTM, SAE, BIS & 50 other standards
TEST SPACE LOADING CAPACITY
Load, max. kg 80
Load per grid kg 20
Possible number of insert grids piece 4
POWER/CONNECTION DATA
Voltage rating V 3/N/PE AC 415V±10% 50Hz other voltages and frequencies optional
Connected Load  kW 28
Sound pressure level dB(A) 76
Total weight kg 1100
Refrigerant System type R404A

R23

(*) Upon customer request, customized sizes are also available.

(*) Append request: Low GWP refrigerants of R449a, R448a, and R508B are available.

(*) The performance data refer to an operating room (ambient) temperature of +26°C, 415 V/50 Hz nominal voltage, without test specimen and without accessories.

(*) Sound Pressure Level: Using a calibrated instrument, the weighted sound pressure level was measured in free-field environments at a height of 1 meter from the floor and a distance of
1 meter from the equipment surface.

Options

(Special accessories on customer request)

Additional Port Holes: Choose from 50mm, 100mm, 125mm, and 150mm port holes to accommodate various connections.
Customizable Floor Loading Capacity: Tailor the chamber's floor loading capacity to meet your specific requirements.
Air Dryer for Low RH Applications: Ensure precise humidity control with an air dryer designed for low RH applications.
Heavy-duty Customized Shelves: Opt for heavy-duty shelves customized to support your specific testing needs.
Water-Cooled Refrigeration Systems: Enhance cooling efficiency with water-cooled refrigeration systems.
Remote Refrigeration/Condensing Unit: Access flexibility and convenience with a remote refrigeration/condensing unit.
Booster Cooling through LN2: Achieve ultra-low temperatures down to -150°C with booster cooling through LN2.

Touchscreen Controller

Experience seamless control with HIACC’s flagship Touchscreen Controller, designed to simplify even the most complex testing procedures. Featuring USB access and Ethernet connectivity, this controller offers unparalleled flexibility.

Here are some of its key features:

HomeScreenProgram_updated
Home
ChamberMenu_updated
Chamber Menu
MIMIC_update
Process MIMIC
Diagnostics_update
Chamber Diagnostics
Alarm_update
Alarm Status

Let's Talk About Your Needs

Select Industry*
Select type of Products*
Select Products*
Select Products*

Download Request

Name*
Contact Number*
Email Id*